MAINTENANCE | Essential Steps for Melt Pressure Transducer Calibration
- Bob

- Sep 5
- 2 min read

Quick Guide: Melt Pressure Transducer Calibration
Audience: Maintenance Leadership
Goal: Get the extruder back up fast with accurate pressure readings
Time Required: ~5 minutes (if no issues)
BEFORE YOU START – Quick Checks
Is the transducer installed and wired properly?
Mounting torque correct (usually 100–200 in-lbs)?
Connector tight?
Wiring verified? (No shorts/opens)
Is the extruder barrel at operating temperature?
Cold barrel = bad readings
Never calibrate a cold transducer
CALIBRATION PROCESS
STEP 1: ZERO CALIBRATION (No Pressure in the System)
You MUST do this before material flow starts.
Power up system – wait 2–3 minutes for sensor to stabilize.
Ensure there is no pressure in the barrel (vent open, no backpressure).
Zero the transducer:
On the amplifier, PLC, or HMI: hit “ZERO” or “TARE”.
If using a digital readout, it should now display 0 psi / 0 bar.
✅ If zero won’t hold, check wiring or sensor damage.
STEP 2: SPAN (OPTIONAL – ONLY IF REQUIRED)
Usually factory-set. Skip unless you’ve replaced the transducer or readings seem off.
Use a simulator or pressure calibrator (if available).
Apply known pressure (e.g., 1000 psi).
Adjust span on the display/controller to match applied value.
✅ If span is off and can’t be adjusted, sensor may be faulty.
STEP 3: VERIFY UNDER LOAD
Start material flow.
Monitor pressure readings under typical extrusion load.
Compare with expected range – if it looks wrong:
Double-check zero
Compare to backup gauge (if installed)
TROUBLESHOOTING – FAST FIXES
Symptom | Possible Cause | Quick Fix |
Stuck at max or min value | Broken sensor or wiring issue | Swap with known good sensor |
Won’t zero | Pressure still in system / damage | Vent barrel / replace sensor |
Fluctuating signal | Ground loop / EMI / loose wires | Check shield, reroute cables |
Reads zero under pressure | Span off or wrong range sensor | Check calibration / specs |
TIPS TO KEEP MOVING
Keep a spare transducer on hand.
Use quick-disconnect cabling if possible.
Don’t calibrate until system is at temp and pressure-free.
Document calibration date & operator for traceability.
Always consult your process management team for FORMAL process steps / these tips are for consideration and to start the conversation!




Comments