Quick Guide: Melt Pressure Transducer Calibration

Audience: Maintenance Leadership

Goal: Get the extruder back up fast with accurate pressure readings

Time Required: ~5 minutes (if no issues)

BEFORE YOU START – Quick Checks

  • Is the transducer installed and wired properly? - Mounting torque correct (usually 100–200 in-lbs)? - Connector tight? - Wiring verified? (No shorts/opens)

  • Is the extruder barrel at operating temperature? - Cold barrel = bad readings - Never calibrate a cold transducer

CALIBRATION PROCESS

STEP 1: ZERO CALIBRATION (No Pressure in the System)

You MUST do this before material flow starts.

  • Power up system – wait 2–3 minutes for sensor to stabilize.

  • Ensure there is no pressure in the barrel (vent open, no backpressure).

  • Zero the transducer: - On the amplifier, PLC, or HMI: hit “ZERO” or “TARE”. - If using a digital readout, it should now display 0 psi / 0 bar.

✅ If zero won’t hold, check wiring or sensor damage.

STEP 2: SPAN (OPTIONAL – ONLY IF REQUIRED)

Usually factory-set. Skip unless you’ve replaced the transducer or readings seem off.

  • Use a simulator or pressure calibrator (if available).

  • Apply known pressure (e.g., 1000 psi).

  • Adjust span on the display/controller to match applied value.

✅ If span is off and can’t be adjusted, sensor may be faulty.

STEP 3: VERIFY UNDER LOAD

  • Start material flow.

  • Monitor pressure readings under typical extrusion load.

  • Compare with expected range – if it looks wrong: - Double-check zero - Compare to backup gauge (if installed)

TROUBLESHOOTING – FAST FIXES

SymptomPossible CauseQuick Fix
Stuck at max or min valueBroken sensor or wiring issueSwap with known good sensor
Won’t zeroPressure still in system / damageVent barrel / replace sensor
Fluctuating signalGround loop / EMI / loose wiresCheck shield, reroute cables
Reads zero under pressureSpan off or wrong range sensorCheck calibration / specs

TIPS TO KEEP MOVING

  • Keep a spare transducer on hand.

  • Use quick-disconnect cabling if possible.

  • Don’t calibrate until system is at temp and pressure-free.

  • Document calibration date & operator for traceability.

Always consult your process management team for FORMAL process steps / these tips are for consideration and to start the conversation!

www.NoNonSensors.com | TechSupport@NoNonSensors.com

Make the next step a confident one.

Share your part number or process requirements. We’ll help you work through the fit.

Talk to an expert